EQUIPMENT BUYING GUIDE · CYNOSURE

Cynosure Elite iQ and Elite iQ PRO: Price, Specs, Skintel and FDA Clearance

Cynosure Elite iQ dual-wavelength laser workstation
Elite iQ. Image: Cynosure. Confirm the model, spot sizes and chiller in your quote.

The Cynosure Elite iQ laser is a two-wavelength platform, with 755 nm alexandrite and 1064 nm Nd:YAG, compatible with a separate cold-air chiller, and the Skintel melanin reader, which Cynosure lists as an accessory; the Elite iQ PRO adds a 30 mm spot and 12 Hz. Cynosure doesn’t publish a price. Here is how to judge a quote without a federal benchmark, what Skintel actually does, what each FDA clearance covers and what to get in writing.

Darryl H.Last updated November 21, 2026Equipment buying guide

Document-based buying analysis, informed by Darryl’s 9+ years in the industry. Not sponsored; NuWays MD has no commercial arrangement with Cynosure. Disclosures.

See price and running costs ⁠↓Elite iQ vs. PRO specs ⁠↓

OUR TAKE

The Elite iQ is Cynosure’s two-wavelength laser platform, with 755 nm alexandrite and 1064 nm Nd:YAG, compatible with a separate cold-air chiller. Its signature feature is Skintel, a handheld melanin reader that the FDA filing describes as a tool for setting a test spot’s starting fluence; Cynosure lists it as an accessory and says to check availability, although both FDA filings describe the system as including it, so confirm it’s in your quote. The Elite iQ was cleared in 2020; the Elite iQ PRO, cleared in 2024, adds a 30 mm spot and repetition rates up to 12 pulses a second. Both wavelengths are cleared for permanent hair reduction, and the 755 nm line names skin types I to VI.

Cynosure doesn’t publish a price, and we found no federal purchase of an Elite iQ, so there’s no public benchmark for this system. For context only, federal buyers paid a median of $107,495 for Candela’s GentleMax Pro Plus and up to $158,301 for Lumenis’s Splendor X. The extra costs to price are a cold-air chiller, if one isn’t included (federal buyers paid $7,000 to $8,345 for Zimmer Cryo 6 chillers, not tied to any laser), and Cynosure service after the warranty. See the price breakdown.

Strengths

  • Alexandrite and Nd:YAG in one platform
  • Skintel reader suggests a test spot’s starting fluence
  • Cold-air chiller compatible, so no cryogen canisters
  • PRO adds a 30 mm spot and up to 12 Hz

Watch for

  • No federal purchase to benchmark a quote
  • The chiller may be a separate purchase
  • No published trial of the Elite iQ or Skintel
  • A 30 A, 208 to 240 V circuit, per Cynosure

See price and running costs ⁠↓

How much does a Cynosure Elite iQ cost?

PRICE CHECK · NOVEMBER 20, 2026

New
Not published by Cynosure. Quotes depend on the model (Elite iQ or PRO), wavelengths, spot sizes and the chiller.
Government orders
None found for the Elite iQ. One for maintenance of an earlier Cynosure Elite: $18,900 for two years, about $9,450 a year (Navy, 2017 to 2019).
Cooling
Standalone Zimmer Cryo 6 cold-air chiller orders, not tied to any laser: $7,000 to $8,345 (2016 to 2026). The filing names no chiller model.
Context
Other two-wavelength systems’ federal orders: GentleMax Pro Plus median $107,495; Splendor X $102,900 to $158,301. Not Elite iQ prices.

Without a public price, your best check is a quote from at least one competitor, configured the same way. We searched federal award records from October 2014 onward and found no Elite iQ purchase; the only Cynosure Elite record in that window was a 2017 Navy maintenance order for an earlier Elite. The Air Force bought an earlier Elite MPX in 2009 for $114,885, too old to benchmark today’s system. Cynosure also sells an alexandrite-only workstation, according to the Elite iQ’s FDA filing, so make sure every quote you compare includes both wavelengths if you need them.

What an Elite iQ quote should spell out
Quote itemWhy it changes the costA comparable answer includes
Model and wave­lengthsElite iQ or Elite iQ PRO, and dual-wavelength or the alexan­drite-only work­station.Exact model, wave­lengths and serial-numbered components
Spot sizesThe PRO’s 30 mm spot and 12 Hz rate are the features to price separately.Every hand­piece and spot included, and the price of any that aren’t
ChillerThe FDA filing lists the Elite iQ as compatible with a cold-air chiller; federal buyers paid $7,000 to $8,345 for Zimmer Cryo 6 chillers, not tied to any laser.Whether a chiller is included, its model and its service terms
SkintelCyno­sure lists Skintel as an accessory and says availability varies.Skintel included, plus any calibration or replacement cost
Service after warrantyTwo systems to keep running: the laser and the chiller.Annual price from year two, coverage, response time and loaners

Is a used Elite iQ cheaper?

The asking price is lower, but the full cost may not be. In Darryl’s experience owning and managing a med spa, a used system can need a manufacturer recertification or inspection fee before the maker will service it. Get in writing whether Cynosure will service the serial number, which handpieces and chiller come with it, the service history, and whether training and a warranty transfer.

Elite iQ vs. Elite iQ PRO: specifications

The PRO keeps the Elite iQ’s wavelengths, fluence ceiling and Skintel reader, and adds a 30 mm spot and a faster maximum repetition rate. Cynosure’s PRO page says the PRO is made by El.En. S.p.A., the Italian company named as applicant on DEKA’s Motus filings (K162886), and distributed exclusively by Cynosure, and it claims 50% faster treatment, up to 80% more power than the previous generation and new 2 and 3 ms pulses. The FDA filing lists the same 600 J/cm² maximum fluence, 0.5 to 300 ms pulse range and 5,500 VA input as the original Elite iQ, so ask Cynosure to show what “faster” and “more power” mean at the spot sizes you’ll use.

Elite iQ and Elite iQ PRO specifications
SpecificationElite iQK193426, 2020Elite iQ PROK240396, 2024
Wave­lengths755 nm alexan­drite and 1064 nm Nd:YAGThe same
Spot sizes2.5, 5, 7, 10, 12, 15, 18, 20, 22 and 24 mmThe same, plus 30 mm
Maximum fluence600 J/cm²600 J/cm²
Pulse duration0.5 to 300 ms0.5 to 300 ms
Repetition rate0.5 to 10 Hz0.5 to 12 Hz
Skin coolingCompatible with a cold-air chillerThe same
Power208 to 240 V, 5,500 VA, single phase; Cyno­sure calls it a 30 A systemThe same
Skintel readerMeasures a 7 x 11 mm area; melanin index 0 to 99, accurate to ±5; under one secondThe same

From FDA 510(k) summaries K193426 and K240396, read November 20, 2026. Spots are interchangeable handpieces with electronic spot recognition. The filings don’t list a maximum pulse energy, so ask Cynosure for the fluence range at the 24 and 30 mm spots.

Area each pulse covers, by spot

Our arithmetic from the spot diameters in each FDA filing.

Elite iQ and PROAdded on the PRO
18 mmBoth models
2.5 cm²
24 mmLargest on the Elite iQ
4.5 cm²
30 mmElite iQ PRO only
7.1 cm²
0510 cm²

What Skintel does, and what it doesn’t

The FDA filing describes Skintel as “an objective measurement tool for examining skin melanin content for determining and setting a test spot starting fluence.” Alongside it, the Elite iQ has a hair removal guidance mode: the provider enters the spot size, skin type and hair type (coarse, medium or fine), and the system suggests a starting fluence and pulse width for a test spot. The filing calls those settings conservative and says users can always change them. Cynosure checked them for safety in a study on the earlier Elite+ laser, not the Elite iQ, examining patients 48 to 72 hours after treatment for side effects; the filing lists it under bench testing, and it didn’t measure hair reduction or test Skintel itself. The study was run by E. Victor Ross, MD, who also appears in Cynosure’s Elite iQ marketing with a quote and a whitepaper.

So Skintel lets a provider start a test spot from a measurement rather than by eye, which Cynosure says helps with tanned skin; the filing aims the guidance mode at less experienced users. It doesn’t replace the test spot, the provider’s judgment or training on darker skin. Cynosure calls it the industry’s first FDA-cleared melanin reader, and the Elite iQ filing compares it with the reader on Cynosure’s Vectus, cleared in 2012. We found no published trial showing that it changes results or complication rates.

MAKE THE QUOTES COMPARABLE

No public price? Make the competitor quote do the work.

Put an Elite iQ quote next to a GentleMax Pro Plus or Splendor X quote: spot sizes, cooling, service after year one and financing, side by side.

Free, works in your browser, no account needed.
Preview of the NuWays MD equipment quote comparison tool
Free equipment quote comparison tool

What is the Elite iQ FDA cleared for?

The FDA cleared the Elite iQ through a 510(k), K193426, on March 9, 2020, and the Elite iQ PRO through K240396 on March 8, 2024. A 510(k) clearance means the FDA found each substantially equivalent to devices already sold, including Cynosure’s Elite+ for the Elite iQ; the PRO’s predicates were the Elite iQ and DEKA’s AGAIN PRO, whose 30 mm handpiece the PRO shares. It isn’t a ranking. The Elite iQ filing lists no clinical tests; it relied on bench, safety and electrical testing.

Elite iQ FDA indications
Wave­lengthHair indi­cationsOther indi­cations
755 nm alexan­driteStable long-term or perma­nent hair reduc­tion, on skin types I to VI in­clud­ing tanned skinVascular lesions, benign pig­mented lesions and wrinkles
1064 nm Nd:YAGStable long-term or perma­nent hair reduc­tion, and pseudo­fol­lic­ulitis barbaeVascular, pig­mented and other benign skin lesions, and wrinkles

From the indications in K193426 and K240396, read November 20, 2026. “Permanent hair reduction” means a long-term, stable reduction in the number of hairs regrowing at 6, 9 or 12 months after a treatment regime, not that every hair is gone.

As with every two-wavelength system, the label isn’t a promise of equal safety on every skin. Your medical director should set which wavelength you use on the darkest skin types, and Cynosure’s own page lists blistering, burns and lightening or darkening of the skin among possible side effects.

What does the research show for the Elite iQ?

We found no published trial of the Elite iQ, the Elite iQ PRO or Skintel. The evidence behind it is the evidence for alexandrite and Nd:YAG lasers in general.

  • In a small randomized, assessor-blinded trial on the legs of people with skin types III and IV, alexandrite and Nd:YAG reduced hair by 74% to 84% at 18 months, with no significant difference between them, though only 15 of 20 people finished; combining both wavelengths added pain and hyperpigmentation (Davoudi and colleagues).
  • A 2020 meta-analysis in skin types III to VI found lasers and IPL broadly similar, with alexandrite ahead of IPL on hair counts and Nd:YAG linked with less hyperpigmentation than IPL (Dorgham and Dorgham).
  • Cynosure’s page quotes a provider who treats a full body “in approximately half the time.” That is a testimonial, not a measured comparison.

Mayo Clinic says most people need 4 to 8 treatments and lists skin color changes, blistering and, rarely, thicker hair growth near the treated area among the risks. Review those risks with your medical director when you set up consent forms, whichever laser you buy.

How many visits pay for an Elite iQ?

With no public price, use your quote. In the worked example in our laser hair removal machines guide, each $100,000 financed over five years at 10% APR costs about $2,125 a month, which takes about 14.4 full-leg visits at $250 or 37.3 underarm visits at $100, after assumed provider time, supplies, card fees and the cost of finding the patient.

PER $100,000 FINANCED

$2,125

A month over 60 months at 10% APR

FULL-LEG VISITS A MONTH

14.4

At $147.50 left each; illustrative

ADD A CHILLER

$7,000+

If it isn’t in the quote; federal orders $7,000 to $8,345

Illustrative: 100% financed, no fees, before tax, installation and service. Visit prices are one practice’s standard single-visit prices (Central Minnesota Dermatology); the costs are our assumptions to replace with your own. Covering the payment isn’t breaking even: rent, other staff and your own pay come on top. Not a quote or a forecast.

What should you compare with the Elite iQ?

Price at least one other two-wavelength system, ideally one with a public price record, so the Elite iQ quote has something to stand against.

Candela GentleMax Pro Plus laser hair removal workstation

Alexandrite + Nd:YAG, cryogen cooled

Candela GentleMax Pro Plus

Candela; FDA cleared May 2020 (K201111)

Both wavelengths with cryogen spray cooling, or an air-cooled option, and spots to 26 mm. Federal buyers paid $92,982 to $123,082 across 16 orders; the lowest may exclude the large-spot kit, which buyers ordered separately for $24,725.

Lumenis Splendor X laser hair removal system

Square spots and a blend mode

Lumenis Splendor X

Sold by Lumenis; FDA filings by Bios S.R.L.

Square spots to 27 mm, round to 30 mm, and an Alex+Nd:YAG configuration, which runs the blend, with a permanent-reduction indication. Federal orders ran $102,900 to $158,301.

DEKA Motus PRO alexandrite and Nd:YAG laser with its touchscreen

Alexandrite + Nd:YAG, in motion

DEKA Motus PRO

DEKA; MOTUS PRO family cleared 2024 (K241459)

Both wavelengths, stamp spots to 24 mm, a 12 Hz system maximum, and the Moveo in-motion technique, which scored far lower for pain than standard pulses in a pilot on the alexandrite-only Motus AX, with lower average hair reduction, a difference the authors reported as not significant.

Images: Candela, Lumenis, DEKA. Compare all eight systems in our laser hair removal machines guide.

What should the demo and written quote prove?

  1. The exact system: Elite iQ or PRO, both wavelengths or alexandrite only, every spot size, whether Skintel is included, and the 510(k) number on the unit.
  2. Skintel in use: readings on your lightest and darkest patients, the suggested starting settings, and how your provider adjusts from the test spot.
  3. A timed large-area pass: legs or a back at 24 or 30 mm and the repetition rate you’d use, by the person who will treat your patients.
  4. The chiller: included or not, its model, its service plan and who repairs it.
  5. Service after the warranty: the annual price from year two for the laser and the chiller, response time and loaners.
  6. Training and resale: the cost to train a new hire, and whether service and software transfer if you sell.

Ask for references from practices with a patient mix like yours. Already bought one? Our paid consulting covers device launches.

Elite iQ buying questions

How much does a Cynosure Elite iQ cost?

Cynosure doesn’t publish a price, and we found no federal purchase of an Elite iQ to benchmark it. Get an itemized quote that names the model, wavelengths, spot sizes and chiller, and compare it with a quote for another two-wavelength system, such as the GentleMax Pro Plus, whose federal orders ran $92,982 to $123,082, the lowest possibly without its large-spot kit.

What is the difference between the Elite iQ and the Elite iQ PRO?

The PRO, cleared by the FDA in 2024, adds a 30 mm spot and raises the maximum repetition rate from 10 to 12 pulses a second. The wavelengths, 600 J/cm² maximum fluence, pulse range, power requirement and Skintel reader are the same, according to the two filings, though the PRO’s software updated the fluence levels for each handpiece, within the Elite iQ’s cleared ranges, to support the 30 mm spot.

What is Skintel on the Elite iQ?

Skintel is a handheld melanin reader, listed by Cynosure as an accessory whose availability you should confirm. The FDA filing describes it as a tool for measuring skin melanin to set a test spot’s starting fluence. It reads a 7 x 11 mm area in under a second. It doesn’t replace the test spot or the provider’s judgment, and we found no published trial of its effect on results.

Does the Elite iQ need a chiller?

Not built in: its FDA filing lists the Elite iQ as compatible with a separate cold-air chiller, which it uses in place of a cryogen spray. Ask whether the quote includes one; federal buyers have paid $7,000 to $8,345 for a Zimmer Cryo 6 chiller.

Can the Elite iQ treat dark skin?

Its FDA filing lists the 755 nm wavelength for skin types I to VI, and 1064 nm Nd:YAG is the wavelength usually chosen for darker skin. We found no published Elite iQ results by skin type, so ask for test spots on your darkest skin types and have your medical director set the protocol; safety depends on settings, cooling and training.

Elite iQ or GentleMax Pro Plus: which is better?

We found no published trial comparing them. Both offer 755 nm and 1064 nm. The Elite iQ is chiller compatible and offers the Skintel reader; the PRO has a 30 mm spot. The GentleMax Pro Plus uses cryogen spray cooling or an air-cooled option, goes to 26 mm and has a public federal price record. Compare timed demos and full quotes, including the chiller or canisters.

Sources and editorial notes

Last updated November 21, 2026. Specifications, FDA wording, federal records and maker claims were checked November 20, 2026. Prepared by Darryl, drawing on 9+ years of med spa experience (he owned and managed a practice for 7+ years, then helped build another), with source-based research. Clinical review pending. Darryl hasn’t owned an Elite iQ; his hands-on hair removal experience is with a handpiece on an Alma Harmony XL Pro. Payment examples are illustrative. NuWays MD is an education and referral resource, not a medical provider. This guide is general business education based on public records, not legal, regulatory or medical advice: confirm indications and advertising wording with the manufacturer and your own counsel, and treatment settings with your medical director. Product names are trademarks of their makers.

View all 17 sources
  1. FDA K193426: Elite iQ (2020): indications, specifications, the Skintel reader, the hair removal guidance mode and its validation on the Elite+, and no clinical tests submitted.
  2. FDA K240396: Elite iQ PRO (2024): the 30 mm spot, 0.5 to 12 Hz, power and Skintel specifications.
  3. FDA: Premarket notification 510(k): what a 510(k) clearance is.
  4. Cynosure: Elite iQ: the dual-wavelength platform, Skintel as the “industry’s first FDA-cleared melanin reader,” a provider testimonial, and the side effects Cynosure lists.
  5. USAspending.gov: searched November 20, 2026 for “Elite iQ,” “Cynosure Elite” and related terms; no Elite iQ system purchase found. Navy order N6890817P0155, “Cynosure Elite laser maintenance,” $18,900 for September 2017 to September 2019.
  6. Cynosure: Elite iQ Pro: the 30 mm spot, Cynosure’s “50% faster,” “up to 80% more power” and “new 2 and 3ms” claims, and the footnote that the PRO is “manufactured by El.En. S.p.A and distributed exclusively by Cynosure.”
  7. USAspending: Air Force order FA700009P0305: “Cynosure Elite MPX laser,” $114,885, August 2009.
  8. USAspending: Air Force order FA301018P0109 and Army order W91YTZ18P0233: Zimmer Cryo 6 chillers at $8,345 and $7,000 (2018), the top and bottom of seven chiller orders from 2016 to 2026.
  9. USAspending: DLA order SPE2D524F512T: GentleMax Pro Plus, $123,082.13, 2024, for context.
  10. USAspending: DLA order SPE2D622F87VQ: Splendor X, $158,301, 2022, for context.
  11. Davoudi and colleagues, Archives of Dermatology, 2008: alexandrite, Nd:YAG and the combination on the legs. Free abstract on PubMed.
  12. Dorgham and Dorgham, JEADV, 2020: lasers and IPL in skin types III to VI. Free abstract on PubMed.
  13. Mayo Clinic: Laser hair removal (July 21, 2026): treatment counts and risks.
  14. FDA K201111: GentleMax Pro Plus and K241459: MOTUS PRO family: the alternatives’ specifications.
  15. Central Minnesota Dermatology: the $100 underarm and $250 full-leg single-visit prices in the example, from a menu headed “DiolazeXL Laser Hair Reduction”; read November 21, 2026.
  16. NuWays MD: Laser hair removal machines: the full worked example and the other systems’ records. It also backs the facts on this page’s comparison cards; each system’s own guide lists its sources in full.
  17. Bonan, Troiano and Verdelli, Dermatologic Therapy, 2020: the Moveo pilot on the Motus AX, for the alternative. Free abstract on PubMed.

Useful next reads

BEFORE YOU SIGN

Know what your laser quote leaves out.

Get the free practice decision kit: a quote-review checklist that goes past the sticker price, plus the planning questions and tools owners use before they commit.

Free. No account needed. Your request isn’t shared with clinics or vendors.

THE KIT INCLUDES

  • A quote-review checklist that goes beyond the sticker price
  • Opening, software and capacity planning questions
  • Direct links to the editable planning tools